Fluke 1529 Chub-E4 Thermometer User Manual


 
8.4.12.8 *TST?
Performs a self-test and reports any errors that are found.
*TST?
0
8.4.12.9 STATus:MEASure?
Reads and clears the Measurement Status Register.
STAT:MEAS?
1
The assignment of the bits in this register is as follows:
bit 0: channel 1 measurement complete
bit 1: channel 2 measurement complete
bit 2: channel 3 measurement complete
bit 3: channel 4 measurement complete
bits 4-15: not used
8.4.12.10 STATus:MEASure:CONDition?
Returns the Measurement Status Condition Register.
STAT:MEAS:COND?
1
The assignment of the bits in this register is as follows:
bit 0: channel 1 measurement in process
bit 1: channel 2 measurement in process
bit 2: channel 3 measurement in process
bit 3: channel 4 measurement in process
bits 4-15: not used
8.4.12.11 STATus:MEASure:ENABle?
Returns the Measurement Status Enable Register.
STAT:MEAS:ENAB?
255
8.4.12.12 STATus:MEASure:ENABle <num>
Sets the Measurement Status Enable Register.
123
8 Digital Communications Interface
Commands