Fluke 1529 Chub-E4 Thermometer User Manual


 
2000,1,1. Specifying the MAX parameter returns a value of 2099,12,31. Spec
-
ifying the DEF parameter returns a value of 2000,1,1.
8.4.5.3 CALibrate<chn>:DATE:CALibrate
(<year>,<month>,<day>)|MIN|MAX|DEF
Sets the calibration date for the specified channel.
CAL1:DATE:CAL 2000,8,29
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. The <year>, <month>, and <day> parame
-
ters must all be supplied. The <year> parameter is a four-digit number (2000 to
2099). The <month> parameter is a one or two-digit number (1 to 12). The
<day> parameter is a one or two-digit number (1 to 31). Specifying the MIN,
MAX or DEF parameter sets the calibration date to the minimum, maximum,
or default value respectively.
This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN <pass> command
(see Section 8.4.11, Password Commands).
8.4.5.4 CALibrate<chn>:DATE:DUE? [MIN|MAX|DEF]
Returns the recalibration date for the specified channel.
CAL1:DATE:DUE?
2001,9,21
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. The response is returned in the format,
<year>,<month>,<day>. Specifying the MIN parameter returns a value of
2000,1,1. Specifying the MAX parameter returns a value of 2099,12,31. Spec
-
ifying the DEF parameter returns a value of 2000,1,1.
8.4.5.5 CALibrate<chn>:DATE:DUE (<year>,<month>,<day>)|MIN|MAX|DEF
Sets the recalibration date for the specified channel.
CAL1:DATE:DUE 2001,8,28
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. The <year>, <month>, and <day> parame
-
ters must all be supplied. The <year> parameter is a four-digit number (2000 to
2099). The <month> parameter is a one or two-digit number (1 to 12). The
<day> parameter is a one or two-digit number (1 to 31). Specifying the MIN,
MAX or DEF parameter sets the recalibration date to the minimum, maximum,
or default value respectively.
This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN <pass> command
(see Section 9.4.11, Password Commands).
1529 Chub-E4 Thermometer Readout
User’s Guide
100