Fluke 1529 Chub-E4 Thermometer User Manual


 
The CALCulate suffix, <chn>, specifies the channel (1 to 4). The AVERage
suffix, <n>, specifies the calculation type. See Table 4 on page 126 for the Sta
-
tistical Calculation Types.
8.4.1.3 CALCulate:AVERage<n>:TYPE?
Returns the keyword for the specified field type number.
CALC:AVER2:TYPE
STD
The AVERage suffix, <n>, specifies the calculation type number. See Table 4
on page 126 for the Statistical Calculation Types.
8.4.1.4 FETCh? [<chn>]
Returns the most recent measurement for the specified channel.
FETC? 1
0.0000011
or
1,2,0.0011,mV,14,5,35,2000,9,5
The <chn> parameter specifies the channel (1 to 4). If the <chn> parameter is
omitted, the most recent measurement is returned regardless of the channel. If
simultaneous sampling is on, the measurement from the lowest enabled channel
number is returned. This command does not affect the measurement action.
If the extended format setting is OFF (see Section 9.4.1.5, FORMat:STAMp
<bool>) the response is only the measurement value in units of temperature,
volts, or ohms depending on the selected probe type or temperature unit. With
this format kohms and millivolts are converted to ohms and volts respectively.
If the extended format setting is ON the measurement is returned in a
comma-delimited format as follows.
Measure status (0 if the measurement has been read previously or 1 if this is a
new measurement,
channel number,
measurement value,
scale,
hour,
minute,
seconds,
year,
month,
89
8 Digital Communications Interface
Commands