Press ENTER to continue …
Press the Enter button to confirm and to continue. The display is remains un
-
changed and all enabled channels remain enabled channels. Press the Exit but
-
ton to cancel.
When the measure period is reset to 1 second or greater, the following message
is displayed indicating that the 1529 is returning to normal operation.
FULL ACCURACY.
The DISPLAY FUNCTIONS function of the CHANNEL menu reappears and
the CHANNEL MODE function is available.
Note: Accuracy in the Fast Measurement Mode is affected by ambient condi
-
tions, changes in measured temperature, higher measured temperature, and the
time duration from using the AUTO-CAL function. To reduce the uncertainty
in the Fast Measurement Mode, use the 1529 in a controlled ambient environ
-
ment, in a very stable bath or dry block at lower temperatures, and use the
AUTO-CAL function often.
7.1.2 Enable Channel
The ENABLE CHANNEL function enables and disables measuring of each
channel. If a channel is off, its measurement is not displayed on the top portion
of the display in either scan or simultaneous display modes. When this function
is selected, the bottom portion of the display shows the on/off status of each
configured channel and allows the setting to be changed. Each configured chan-
nel is identified by its channel number.
Use the UD buttons to select the channel to change. The LR buttons are
used to change the setting. Press the Enter button to save the setting. Press and
1529 Chub-E4 Thermometer Readout
User’s Guide
32
1
24.159C
3
24.142C
2
24.634C
4
24.015C
CHAN 1: ON
CHAN 2:
ON
CHAN 3:
ON
CHAN 4:
ON
EXIT
ENTER / MENU UNIT CONTRAST
1529
CHUB E-4 THERMOMETER READOUT
Figure 9 Enable Channel