Fluke 1529 Chub-E4 Thermometer User Manual


 
the specified channel. Specifying the MIN, MAX, or DEF parameter sets the
RJC parameter to the minimum, maximum, or default value respectively.
This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN <pass> command
(see Section 9.4.11, Password Commands).
8.4.6 Display Commands
The display commands are used for querying and setting the display resolution,
lamp brightness, decimal format, window setting, and window fields.
8.4.6.1 DISPlay:RESolution? [MIN|MAX|DEF]
Returns the display resolution setting.
DISP:RES?
2
Specifying the MIN pamameter returns a value of 0. Specifying the MAX pa-
rameter returns a value of 4. Specifying the DEF parameter returns AUT.
8.4.6.2 DISPlay:RESolution <num>|AUT|MIN|MAX|DEF
Sets the display resolution setting.
DISP:RES 3
The <num> parameter specifies the number of decimal places to display. This
parameter can be 0, 1, 2, 3, or 4. Specifying the MIN, MAX or DEF parameter
sets the display resolution to the minimum, maximum, or default value respec-
tively. Specifying the AUT parameter sets the display resolution to automatic.
8.4.6.3 DISPlay:LAMP? [MIN|MAX|DEF]
Returns the lamp brightness setting.
DISP:LAMP?
0
Specifying the MIN parameter returns a value of 0. Specifying the MAX pa
-
rameter returns a value of 3. Specifying the DEF parameter returns a value of 0.
8.4.6.4 DISPlay:LAMP <bool>|<num>|MIN|MAX|DEF
Sets the lamp brightness setting.
DISP:LAMP ON
The lamp can be turned on or off by using either the <bool> parameter (ON or
OFF), the <num> parameter (0, 1, 2 or 3), or the MIN, MAX or DEF parame
-
ters. Specifying the <bool> parameter to ON turns on the lamp. Specifying the
1529 Chub-E4 Thermometer Readout
User’s Guide