Fluke 1529 Chub-E4 Thermometer User Manual


 
8.4.10.7 SYSTem:TIME?
Returns the current system time.
SYST:TIME?
11,43,23
The response is returned in the format, <hour>,<minute>,<second>.
Note: The system time is always returned using 24-hour format regardless of
the current time format setting.
8.4.10.8 SYSTem:TIME <hour>,<minute>,<second>
Sets the current system time to the specified hour, minute, and second.
SYST:TIME 11,43,23
The <hour>, <minute>, and <second> parameters must all be supplied. The
<hour> parameter is a one or two-digit number (0 to 23). The <minute> param
-
eter is a one or two-digit number (0 to 59). The <second> parameter is a one or
two-digit number (0 to 59).
Note: The system time is always set using 24-hour format regardless of the cur-
rent time setting.
8.4.11 Password Commands
The password commands are used for enabling and disabling the password-pro-
tected commands, for querying and setting the probe password-protected state,
and for setting a new password.
8.4.11.1 SYSTem:PASSword:CDISable
Disables password protected commands.
SYST:PASS:CDIS
Sending this command prevents the user from accessing the password-pro
-
tected settings.
8.4.11.2 SYSTem:PASSword:CENable <pass>
Enables password protected commands.
SYST:PASS:CEN 1234
The <pass> parameter is the current system password. This command must be
sent to allow the user to access password-protected settings. After sending this
command, use the SYST:PASS:CEN:STAT? command to verify success.
8.4.11.3 SYSTem:PASSword:CENable:STATe?
Returns the current password protection state.
1529 Chub-E4 Thermometer Readout
User’s Guide
120