By default the password is set to ‘1529’ when the instrument ships from the
factory. We recommend that the user change the default password to protect the
password-protected settings. Note: Keep your password in a secure location
and do not forget it.
To enter the password, press the Enter button. Press the LR buttons to high
-
light the digit to change. Use the UD buttons to scroll through the numbers
0-9. Press the Enter button to save the new setting. Press the Exit button to
cancel and to exit to the menu.
If an incorrect password is entered the display shows the message PASSWORD
INCORRECT and access to the password-protected parameters is denied. Press
the Enter button to return to the menu. If the correct password is entered, the
bottom portion of the display instructs the user to press Enter to reset the set
-
tings to the defaults.
The system reset does the following:
Sets the measure period to 1 second.
Sets channel 1 ON, all others OFF.
Sets the channel mode to SCAN.
Sets moving average to 1.
Sets printing to all ports OFF.
Stops auto logging.
Clears the statistical functions.
Sets units to C.
Sets the resolution to AUTO.
Reset does not affect any of the probe characterization parameters, logged data
or channel calibration parameters.