Configuration
2.10.1 Device reset
The master reset function will reset the device electronics. To perform a device reset:
Performing master reset using a Field Communicator
Performing master reset using AMS Device Manager
1. From the Home screen, select 3: Service Tools.
2. Select 5: Maintenance
3. Select 5: Device Reset
2.10.2 Join status
Viewing join status using a Field Communicator
Viewing join status using AMS Device Manager
To view the join status of the device, perform the following procedure:
1. From the Home screen, select 3: Service Tools.
2. Select 4: Communications.
3. Select 1: Join Status.
Wireless devices join the secure network through a four step process:
Step 1. Network Found
Step 2. Network Security Clearance Granted
Step 3. Network Bandwidth Allocated
Step 4. Network Join Complete
From the HOME screen, enter the fast key sequence
Device Dashboard Fast Keys
3, 5, 5
From the HOME screen, enter the fast key sequence
Device Dashboard Fast Keys
3, 4, 1