Emerson Process Management IB106-340 Oxygen Equipment User Manual


 
Instruction Manual
IB-106-340 Rev. 2.4
April, 2001
Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 5-9
Oxymitter 4000
5-8 DEFINING A TIMED CALIBRATION VIA
HART
Use the following procedure to specify a time
interval (in hours) at which the Oxymitter 4000
will be automatically calibrated.
NOTE
To select a menu item, either use the
up and down arrow keys to scroll to
the menu item and press the right ar-
row key or use the number keypad to
select the menu item number.
To return to a preceding menu, press
the left arrow key.
a. From the DEVICE SETUP screen, select
DETAILED SETUP.
b. From the DETAILED SETUP screen, select
O2 CALIBRATION.
c. From the O2 CALIBRATION screen, select
menu item 6, CAL MODE. Set the CAL
MODE to AUTO.
d. Return to the O2 CALIBRATION screen and
select menu item 7, CAL INTRVL.
e. At the prompt, input a time interval (in
hours) at which an automatic calibration will
occur and press ENTER.