Figure 7-14. Fault 14, Invalid Constant
n. Fault 14, Invalid Constant
The CALIBRATION LED flashes twice,
pauses for three seconds, and repeats
(Figure 7-14).
1. After a calibration has been performed,
the electronics calculates a cell con-
stant value.
2. If the cell constant value is outside of
the range, -4 mV to 10 mV, the alarm
will activate. See paragraph 6-2, Cali-
bration, and verify the last calibration
was performed correctly.
3. Power down the Oxymitter 4000 and
remove it from the stack.
4. Replace the cell per paragraph 6-8,
Cell Replacement.
7