Emerson Process Management IB106-340 Oxygen Equipment User Manual


 
Instruction Manual
IB-106-340 Rev. 2.4
April, 2001
Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 7-11
Oxymitter 4000
DIAGNOSTIC
ALARMS
TEST
POINTS
HEATERT/C
HEATER
O2 CELL
CALIBRATION
CALIBRATION RECOMMENDED
O2 CELL mV +
O2 CELL mv -
HEATERT/C +
HEATERT/C -
INC INC
DEC DEC
HIGH
GAS
LOW
GAS
CAL
TEST GAS +
PROCESS -
% O2
SW2
TP1
J1
TP2
TP3
RED
YEL
GRN
ORG
TP4
TP5
TP6
ON
22220039
Figure 7-8. Fault 8, Low Heater Temp
h. Fault 8, Low Heater Temp
The HEATER LED flashes four times,
pauses for three seconds, and repeats
(Figure 7-8).
1. The low heater temperature alarm is
active when the thermocouple reading
has dropped below 28.6 mV.
2. If the thermocouple reading continues
to ramp downward for one minute and
does not return to the temperature set
point of approximately 29.3 mV, then
an Open Heater fault will be displayed.
3. Power down the electronics. Remove
the electronic assembly per paragraph
6-5b, Electronic Assembly Replace-
ment. Using a multimeter, measure
across the heater connector, J8.
4. If the heater is good, the reading will be
approximately 70 ohms. If the heater is
open, see paragraph 6-7, Heater Strut
Replacement.
7