Emerson Process Management IB106-340 Oxygen Equipment User Manual


 
Instruction Manual
IB-106-340 Rev. 2.4
April, 2001
2-14 Installation Rosemount Analytical Inc. A Division of Emerson Process Management
Oxymitter 4000
TO ELECTRONICS
REF AIR SET
263C152G01
1 FLOWMETER 0.2-2.0 SCFH 771B635H02
2 2" PRESSURE GAGE 0-15 PSIG 275431-006
3 COMBINATION FILTER-REG. 0-30 PSIG 4505C21G01
NOTE: DIMENSIONS ARE IN INCHES WITH
MILLIMETERS IN PARENTHESES.
1
2
3
4.81 (122.17)
FLOW SET
POINT KNOB
0.125-27 NPT FEMALE
OUTLET CONNECTION
1.19
(30.22)
10.0
(254)
REF
DRAIN VALVE
3.12 (79.25) MAX
8.50
(215.90)
MAX
2.0
(50.80)
2 MOUNTING HOLES
3.19 (81.03) LG
THROUGH BODY FOR
0.312 (7.92) DIA BOLTS
1.50
(38.10)
2.250 (57.15)
SCHEMATIC HOOKUP FOR REFERENCE AIR SUPPLY ON OXYMITTER 4000 PROBE HEAD.
OUTLET
0.25-18 NPT FEMALE
INLET CONNECTION
INSTRUMENT AIR SUPPLY
10-225 PSIG MAX PRESSURE
26170035
0.250 OR 6 MM O.D. TUBING
(SUPPLIED BY CUSTOMER)
FITTING (SUPPLIED BY WECO)
TUBE COMPRESSION
0.250 OR 6 MM O.D.
500 VA
SERIAL NO.
TAG NO.
OXYMITTER 4000
WATTS:VOLTS:
FUSE:LINEOUTPUT:
Rosemount Analytical Inc.
Orrville,OH 44667-0901
85-264 VAC 48-62 Hz
TM
800-433-6076
4-20 mA
R
5 Amps
TM
HART
SMART FAMILY
Figure 2-11. Air Set, Plant Air Connection
2-4 PNEUMATIC INSTALLATION (FOR OXY-
MITTER 4000 WITHOUT SPS 4000)
a. Reference Air Package
After the Oxymitter 4000 is installed, con-
nect the reference air set to the Oxymitter
4000. The reference air set should be in-
stalled in accordance with Figure 2-11.
Instrument Air (Reference Air): 10 psig
(68.95 kPag) minimum, 225 psig (1551.38
kPag) maximum at 2 scfh (56.6 L/hr) maxi-
mum; less than 40 parts-per-million total
hydrocarbons. Regulator outlet pressure
should be set at 5 psi (35 kPa). Reference
air can be supplied by the reference air set
of the IMPS 4000.
If using an IMPS 4000, refer to the IMPS
4000 Intelligent Multiprobe Test Gas Se-
quencer Instruction Bulletin for the proper
reference air connections.