7-14 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Oxymitter 4000
DIAGNOSTIC
ALARMS
TEST
POINTS
HEATERT/C
HEATER
O2 CELL
CALIBRATION
CALIBRATION RECOMMENDED
O2 CELL mV +
O2 CELL mv -
HEATERT/C +
HEATERT/C -
INC INC
DEC DEC
HIGH
GAS
LOW
GAS
CAL
TEST GAS +
PROCESS -
% O2
SW2
TP1
J1
TP2
TP3
RED
YEL
GRN
ORG
TP4
TP5
TP6
ON
22220042
Figure 7-11. Fault 11, Bad Cell
k. Fault 11, Bad Cell
The O2 CELL flashes three times, pauses
for three seconds, and repeats (Figure
7-11).
1. The bad cell alarm activates when the
cell exceeds the maximum resistance
value.
2. The cell should be replaced. See para-
graph 6-8, Cell Replacement, for cell
replacement instructions.