Figure 7-10. Fault 10, High Cell mV
j. Fault 10, High Cell mV
The O2 CELL flashes once, pauses for
three seconds, and repeats (Figure 7-10).
1. Using a multimeter, measure across
TP1+ to TP2-.
2. If you measure 1.2 VDC, the cell wires,
either orange or green, have become
detached from the input.
3. One possible cause is connector J1.
The orange or green wire has come
loose from the crimped connection.
4. The platinum pad could also be at fault.
The pad could have broken free from
the back of the cell.
5. Replace heater strut per paragraph
6-7, Heater Strut Replacement. If
necessary, replace the cell flange
assembly per paragraph 6-8, Cell
Replacement.