Figure 7-1. Fault 1, Open Thermocouple
a. Fault 1, Open Thermocouple
The HEATER T/C LED flashes once,
pauses for three seconds, and repeats
(Figure 7-1).
1. Check connector J1. Ensure the con-
nector is properly seated.
2. Using a multimeter, measure TP3+ to
TP4-. If the reading is 1.2 VDC ±0.1
VDC, the thermocouple is open.
3. Remove power. Disconnect J1. Meas-
ure continuity across the red and yel-
low thermocouple leads.
4. The measurement should read ap-
proximately 1 ohm.
5. If the thermocouple is open, see para-
graph 6-7, Heater Strut Replacement.