Emerson Process Management IB106-340 Oxygen Equipment User Manual


 
Instruction Manual
IB-106-340 Rev. 2.4
April, 2001
7-12 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management
Oxymitter 4000
DIAGNOSTIC
ALARMS
TEST
POINTS
HEATERT/C
HEATER
O2 CELL
CALIBRATION
CALIBRATION RECOMMENDED
O2 CELL mV +
O2 CELL mv -
HEATERT/C +
HEATERT/C -
INC INC
DEC DEC
HIGH
GAS
LOW
GAS
CAL
TEST GAS +
PROCESS -
% O2
SW2
TP1
J1
TP2
TP3
RED
YEL
GRN
ORG
TP4
TP5
TP6
ON
22220040
Figure 7-9. Fault 9, High Heater Temp
i. Fault 9, High Heater Temp
The HEATER LED flashes five times,
pauses for three seconds, and repeats
(Figure 7-9).
1. If the thermocouple produces a voltage
in excess of approximately 30.7 mV,
the high heater temp alarm activates.
2. The 4-20 mA signal returns to the de-
fault value (4 or 20 mA).
3. This alarm is self-clearing. When tem-
perature control is restored and the
thermocouple voltage returns to the
normal range, the alarm clears.
4. If the temperature continues to rise, the
next alarm will be the high high heater
temp alarm.