Invacare M61 Mobility Aid User Manual


 
SECTION 3—SAFETY INSPECTION/TROUBLESHOOTING
Part No. 1125075 29 Pronto® Series
M61 Only
Checking Battery Charge Level
The following “Do’s” and “Don’ts” are provided for your convenience and safety.
Batteries have
short driving range
during a single
charge. Battery
Gauge falls off
faster than normal.
Consumer not charging batteries long
enough.
Instruct consumer to charge for 8-10 hours
minimum.
Batteries may be weak. Perform load test or check “Battery Quality Menu”
with the programmer. Refer to electronics manual.
Refer to Field Load Test
on page 30.
Check programming settings. Torque setting and power level setting may be too
high. Refer to electronics manual.
Heavy load on motors. Chairs weight distribution may be offset
(wheelchair may be front loaded).
SYMPTOM PROBABLE CAUSE SOLUTIONS
Wheelchair slows
or stops while
driving and the Seat
Function I indicator
is lit.
Elevating seat is elevated. Return seat to its lowest position. Refer to
Elevating the Seat
on page 37.
Elevating seat sensor is damaged. Examine elevating seat sensor. Replace if necessary.
Refer to Removing/Installing the Actuator Switch
Sensor on page 97.
DO DON’T
Read and understand this manual and any service
information that accompanies a battery and charger
before operating the wheelchair.
Don’t perform any installation or maintenance without
first reading this manual.
Move the wheelchair to a work area before opening
battery box or installing service batteries.
Don’t perform installation or maintenance of batteries
in an area that could be damaged by battery spills.
Recharge as frequently as possible to maintain a high
charge level and extend battery life.
Don’t make it a habit to discharge batteries to the
lowest level.
Follow recommendations in this manual when
selecting a battery or charger.
Don’t use randomly chosen batteries or chargers.
Fully charge new batteries before using. Don’t put new batteries into servcie before charging.
Use a carrying strap to remove, move or install a
battery.
Don’t tip or tilt batteries.
Push battery clamps on the terminals. Spread clamps
wider if necessary.
Don’t tap on clamps and terminals with tools.
Use ONLY a GEL charger for a GEL battery or
“Sealed” battery.
Don’t mismatch your battery and chargers.
SYMPTOM PROBABLE CAUSE SOLUTIONS