Invacare IRC5LX Respiratory Product User Manual


 
SECTION 20—4-WAY VALVE
Platinum™Series 82 Part No 1110538
ƽ WARNING
Ensure that hoses behind control panel DO NOT contact cooling fan after control
panel reinstallation.
20. Plug power cord in and turn concentrator ON (I) to ensure proper operation.
21. Reinstall cabinet. Refer to Removing Cabinet on page 25.
22. If any difficulties are encountered in this installation, contact Invacare Technical
Service at (800) 832-4707.
FIGURE 20.5 Replacing Pilot Valve Poppets and O-Rings
Spade Connectors attach here
(Leave Wire Connectors Intact)
Pilot Valve
Stem
Yoke
New
Label
Locknut
Coil
Pilot Valve
Stem
Valve Body
(Reference)
Manifold
Tapered
End
O-Ring
Slotted
Poppet
Manifold
Opening
NOTE: DO NOT
disassemble.
Air Flow Passage
Manifold
Opening
O-Ring
Platinum 5 and XL Models
Platinum 10 Models
DETAIL “A”
Locknut
(hidden
from view)
Coil
Pilot Valve
Stem
Valve Body
(Reference)
Tapered
End
O-Ring
Slotted
Poppet
Manifold Opening
NOTE: DO NOT disassemble.