Invacare IRC5LX Respiratory Product User Manual


 
TYPICAL PRODUCT PARAMETERS
Platinum™Series 10 Part No 1110538
WEIGHT: Platinum 5 and XL: 52 lbs ± 2 lbs (23.6 kg ± 1 kg).
Platinum 10: 53 lbs ± 2 lbs (24.0 kg ± 1 kg).
SHIPPING WEIGHT: Platinum 5 and XL: 58 lbs ± 2 lbs (26.3 kg ± 1 kg).
Platinum 10: 60 lbs ± 2 lbs (27.2 kg ± 1 kg).
OPERATING AMBIENT
TEMPERATURE:
50°F - 95°F (10°C - 35°C) at 20-60% relative humidity.
EXHAUST: Platinum 5 and XL: Less than Ambient + 35°F (+19°C).
Platinum 10: Less than Ambient + 45°F (+25°C).
OXYGEN OUTPUT:
Platinum 5 and XL: Less than Ambient + 4°F (+2°C).
Platinum 10: Less than Ambient + 8°F (+4.4°C).
CABINET: Impact Resistant flame-retardant plastic cabinet that conforms
to UL 94-V0.
REGULATORY LISTING: Double Insulated Product ETL certified complying with UL 1097
ETL certified to CSA C22.2 No. 68.
ELECTRICAL: No extension cords.
PLACEMENT: No closer than three inches from any wall, furniture, draperies,
or similar surfaces.
TUBING: Platinum 5 and XL: 7 ft cannula with a maximum 50 ft of
Crush-Proof Tubing (DO NOT pinch).
Platinum 10: Recommended use up to 50 ft high flow tubing
with high flow cannula at all flow rates.
RELATIVE HUMIDITY: 20 to 60%.
TIME OF OPERATION: Up to 24 hours per day.
RECOMMENDED STORAGE AND
SHIPPING TEMPERATURE:
-20°F to 150°F (-29°C to 65°C), at 15-95% relative humidity.
LABEL
SYMBOL
STATUS SENSO
2
MODELS INDICATOR LIGHTS (LED)
SYSTEM OKAY
O
2
over 85% (± 2%)
GREEN Indicator Light
O
2
Between 73% (± 3%) to 85%
(±2%)
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor Failure
Call a qualified technician.
SYSTEM FAILURE
O
2
Below 73% (±3%)
RED Indicator light
Continuous Audible Alarm Sieve-GARD™Compressor Shutdown
Call a qualified technician
LABEL
SYMBOL
STATUS LX MODEL INDICATOR LIGHTS (LED)
SYSTEM OKAY GREEN Indicator Light
SYSTEM FAILURE RED Indicator light
Continuous Audible Alarm Sieve-GARD™ Compressor Shutdown
Call a qualified technician
O
2
I/O