Invacare IRC5LX Respiratory Product User Manual


 
SECTION 7—COMPRESSOR
Platinum™Series 38 Part No 1110538
Rubber Mallet
Phillips screwdriver attachment for torque wrench (for retainer screws)
CAUTION
DO NOT lubricate or use oil on any moving parts. The compressor eccentric uses a
precision sealed bearing. Therefore, additional lubrication is not necessary.
Removing Compressor and Compressor Heads
Compressor
1. Unplug concentrator and carefully remove compressor from concentrator. Refer to
Replacing Compressor Assembly on page 33.
Compressor Head Plate
NOTE: For this procedure, refer to FIGURE 7.4.
1. Clean loose dirt from the outside of the
compressor.
2. Use the Torx T-25 torque wrench and
loosen the eight head screws.
3. Note head placement before removing
compressor heads.
4. Remove the compressor head plate.
Valve Plates
NOTE: For this procedure, refer to FIGURE 7
.4.
1. Note the position of the plates before
removing.
2. Lift off the valve plate on one side.
3. Repeat for the other side.
NOTE: Note the position of the tab on valve
plate for reassembly.
NOTE: To avoid confusion, ONLY service one
(1) end of the compressor at a time.
FIGURE 7.4 Removing Compressor and
Compressor Heads - Rebuilding the
Compressor - Compressor - Compressor
Head Plate - Valve Plates
Head
Screws
Head Screws
Compressor
Head Plate
Valve
Plate
Valve
Plate
Compressor