Invacare IRC5LX Respiratory Product User Manual


 
SECTION 4—TROUBLESHOOTING
Part No 1110538 17 Platinum™Series
SECTION 4—TROUBLESHOOTING
SYMPTOM PROBABLE CAUSE SOLUTION
NORMAL OPERATION:
Internal Status Indicators:
RED: Off
GREEN: Off
Unit plugged in, power
switch on. Single beep on
start up.
No Problems System Okay.
POWER LOSS:
Internal Status Indicators:
RED: Off
GREEN: Off
Unit unplugged, power
switch on, alarm off.
No Problems. Battery Free Circuit drained. Plug in cord
and turn power switch ON (I) to recharge.
POWER LOSS:
Internal Status Indicators:
RED: Off
GREEN: Off
Unit plugged in, power
switch, alarm off, battery
Free circuit drained.
No power at outlet. Check electrical outlet with a table lamp or
voltmeter set on 50-200 VAC scale. If
outlet isn't working, check protective
device in home's electrical panel or consult
an electrician. Also ensure that unit is
properly plugged in. DO NOT use
extension cords. Move to another outlet
or circuit.
Power cord:
a. Frayed.
b. Broken or damaged spade.
c. Spade connector from power cord
loose or disconnected (inside back
of unit).
a. Reattach cord.
b. Replace power cord connectors on
plug.
c. Reattach connector.