SECTION 6—PREVENTIVE MAINTENANCE
Platinum™Series 30 Part No 1110538
3. Remove the two mounting screws from rear of sound box.
4. Disconnect and remove the manifold exhaust tubing from the barbed fittings on top
of the exhaust canister.
5. Unscrew the exhaust muffler located on the exhaust canister by turning
counterclockwise.
6. Inspect exhaust canister to ensure the barbed fittings are clear and do one of the
following:
Replace exhaust canister and muffler if plugged or restricted.
Install only new exhaust muffler if exhaust canister is in good condition.
7. Install new/re-install existing exhaust canister by reversing STEPS 3-5.
8. Reinstall cabinet. Refer to Removing Cabinet on page 25.
FIGURE 6.5 Replacing the Exhaust Canister/Exhaust Muffler (Platinum 5 Models Manufactured
before 7/24/03)
Exhaust Canister
Exhaust Muffler
Barbed Fittings
Manifold Exhaust Tubing
Mounting Screw(s)
Rear of Sound Box