Invacare IRC5LX Respiratory Product User Manual


 
SECTION 7—COMPRESSOR
Platinum™Series 36 Part No 1110538
4. Disconnect the spade connectors from the capacitor terminals (not shown).
5. Remove the screw that secures the capacitor clamp to the soundbox brace.
6. Remove the capacitor from the concentrator base assembly.
7. Install new capacitor by reversing STEPS 3-6.
8. Re-install the cabinet. Refer to Removing Cabinet on page 25.
10-Liter Model (Platinum 10)
NOTE: For this procedure, refer to Detail “B” in FIGURE 7.3.
1. Unplug the unit.
2. Remove the cabinet. Refer to Removing Cabinet on page 25.
3. Cut the two tie-wraps that secure the capacitor to the soundbox brace.
4. Rotate capacitor to horizontal position.
5. Remove protective boot to reveal capacitor wires.
6. Disconnect the spade connectors from the capacitor terminals (not shown).
7. Remove the capacitor from the concentrator base assembly.
8. Install new capacitor by reversing STEPS 3-7.
9. Re-install the cabinet. Refer to Removing Cabinet on page 25.
FIGURE 7.3 Replacing Capacitor
Soundbox
Brace
Protective
Boot
Tie-Wraps
Capacitor
DETAIL “B” - PLATINUM 10 MODELS
Capacitor
Screw
Soundbox
Brace
Protective
Boot
Capacitor
Clamp
DETAIL “A” - PLATINUM 5 AND
PLATINUM XL MODELS