Invacare IRC5LX Respiratory Product User Manual


 
SECTION 5—CABINET
Part No 1110538 25 Platinum™Series
SECTION 5—CABINET
Removing Cabinet
ƽ DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical out-
let before servicing.
NOTE: For this procedure, refer to FIGURE 5.1.
1. Unplug unit.
2. Remove the four mounting screws that secure cabinet assembly to the base assembly.
3. Lift the cabinet straight up.
NOTE: When required, vacuum inside of the cabinet and exposed foam insulation.
4. To re-install cabinet, reverse STEPS 2-3.
L
FIGURE 5.1 Removing Cabinet
Cabinet Assembly
Mounting
Screws
Base Assembly
Mounting
Screws