Invacare IRC5LX Respiratory Product User Manual


 
SECTION 6—PREVENTIVE MAINTENANCE
Platinum™Series 26 Part No 1110538
SECTION 6—PREVENTIVE
MAINTENANCE
ƽ DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical
outlet before servicing.
NOTE: The Platinum Concentrators are specifically designed to minimize routine preventive
maintenance at intervals of once per year. In places with high dust or soot levels, maintenance may
need to be performed more often. The following MUST be performed at a minimum of one year in
service to assure years of additional reliability. Refer to Troubleshooting on page 17 for plugged
filter symptoms. Only qualified technicians should perform preventive maintenance on the
concentrator. Power should be disconnected before beginning preventive maintenance on the
Invacare concentrator.
Cleaning the Cabinet Filter
NOTE: For this procedure, refer to FIGURE 6.1.
NOTE: Perform this procedure as needed depending upon the environment the concentrator is
used in.
NOTE: There are two cabinet filters, one located on each side of the cabinet.
1. Remove each filter and clean at least once a week depending on environmental
conditions.
NOTE: Environmental conditions that may
require more frequent cleaning of the filters
include but are not limited to: high dust, air
pollutants, etc.
2. Clean the cabinet filters with a vacuum
cleaner or wash in warm soapy water
and rinse thoroughly.
3. Dry the filters thoroughly before
reinstallation.
CAUTION
DO NOT operate the concentrator
without the filters installed.
FIGURE 6.1 Cleaning the Cabinet Filter
Cabinet Filter
(located on both
sides of
concentrator)
Cabinet Filter