Tektronix WFM 601M Wheelchair User Manual


 
Appendix A: Specifications
A–4
WFM 601A, WFM 601E & WFM 601M User Manual
Table A–1: Electrical specifications (cont.)
Characteristic Reference informationPerformance requirement
Dynamic Range 10 UI p-p
Deflection Factor
Horizontal Four modes available:
1 Line, 2 Line
1 Field, 2 Field
Vertical
X1 1 UI/div $5%
X5 0.2 UI/div $10%
X10 0.1 UI/div $20%
–3 dB Bandwidths
Bandwidth Selection 10 Hz to 5 MHz
1 kHz to 5 MHz
10 kHz to 5 MHz
100 kHz to 5 MHz
Jitter Output (WFM 601M Only)
120 mV/UI $10% into 75 W load.
Signal Level & Cable Length Detector
(WFM 601E and WFM 601M Only)
Cable Length
Accuracy 10 meters $10% of reading (Belden 8281 cable).
Resolution 5 meters
Dynamic Range 0 to 400 meters (Belden 8281 cable)
Source Signal Level
Accuracy $0.5 dB for cable lengths from 0–100 meters
(Belden 8281)
Resolution 0.25 dB
Dynamic Range $3 dB
Cable Types Supported cable types:
Belden 8281
Belden 1694A
Belden 1505A
Video Error Detection
Type Active picture and full field with field rate resolution.
Uses EDH (Error Detection and Handling) system
based on CRC check-word. Complies with SMPTE
RP165.
Reporting Means Front-panel ALARM indicator, rear-panel TTL line,
and CRT readout.
Error Statistics Asynchronous errored seconds. Active picture and
full field statistics are separately compiled.