Emerson Process Management IM-106-340 Oxygen Equipment User Manual


 
Oxymitter 4000
1-16
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
SPECIFICATIONS
Oxymitter Specifications
O
2
Range
Standard 0 to 10% O
2
,
0 to 25% O
2
,
0 to 40% O
2
(via HART)
Accuracy ±0.75% of reading or 0.05% O
2
, whichever is greater
System Response to Calibration
Gas
Initial – less than 3 seconds, T90 – less than 8
seconds
Temperature Limits
Process 32° to 1300°F (0° to 704°C) up to 2400°F (1300°C)
with optional accessories
Electronics Housing -40° to 158°F (-40° to 70°C) ambient
Electronics Package -40° to 185°F (-40° to 85°C) [Operating temperature
of electronics inside of instrument housing, as
measured by a HART communicator, Rosemount
Analytical Asset Management Solutions software.]
Local Operator Interface -40° to 158°F (-40° to 70°C), [above 158°F (70°C) the
infrared keypad will cease to function, but the
Oxymitter 4000 will continue to operate properly.]
Probe Lengths 18 in. (457 mm) 12 ft (3,66 m)
3 ft (0,91 m) 15 ft (4,57 m)
6 ft (1,83 m) 18 ft (5,49 m)
9 ft (2,74 m)
Mounting and Mounting Position Vertical or horizontal;
a spool piece, (P/N 3D39761G02), is available to
offset transmitter housing from hot ductwork.
Materials
Probe Wetted or welded parts - 316L stainless steel (SS)
Non-wetted parts - 304 SS, low-copper aluminum
Electronics Enclosure Low-copper aluminum
Calibration Manual, semi-automatic, or automatic
Calibration Gas Mixtures
Recommended
0.4% O
2
, Balance N
2
8% O
2
, Balance N
2
Calibration Gas Flow 2.5 l/m (5 scfh)
Reference Air 2 scfh (1 l/m), clean, dry, instrument-quality air
(20.95% O
2
), regulated to 34 kPa (5 psi)
Electronics NEMA 4X, IP66 with fitting and pipe on reference
exhaust port to clear dry atmosphere
Electric Noise EN 61326-1, Class A
Certifications General Purpose
Line Voltage 90-250 VAC, 48/62 Hz. No configuration necessary.
3/4 in. -14 NPT conduit port
Table continued on next page
APPROVED
US
C