Emerson Process Management IM-106-340 Oxygen Equipment User Manual


 
Oxymitter 4000
9-16
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
Figure 9-9. Probe to Probe Head
Assembly - Remote Electronics
Only
Cell Replacement
This paragraph covers oxygen sensing cell replacement. Do not attempt to
replace the cell until all other possibilities for poor performance have been
considered. If cell replacement is needed, order the cell replacement kit
(Table 10-1). Refer to Figure 9-3 or Figure 9-4 to view the component parts of
the Oxymitter 4000.
The cell replacement kit (Figure 9-10) contains a cell and flange assembly,
corrugated seal, setscrews, socket head cap screws, and anti-seize
compound. The items are carefully packaged to preserve precise surface
finishes. Do not remove items from the packaging until they are ready to be
used. Spanner wrenches and hex wrenches needed for this procedure are
part of an available special tools kit (Table 10-1 and Figure 10-2).
Probe Signal
Wire Connector
Probe Heater
Wire Connector
4 Wire
2 Wire
Housing
O-Ring
Probe
37260058
Adapter
Board