Emerson Process Management IM-106-340 Oxygen Equipment User Manual


 
Oxymitter 4000
viii
Instruction Manual
IM-106-340, Rev. 4.0
May 2006
QUICK START GUIDE
FOR OXYMITTER 4000
SYSTEMS
Before using the Quick Start Guide, please read "WHAT YOU NEED TO
KNOW BEFORE INSTALLING AND WIRING A ROSEMOUNT
ANALYTICAL OXYMITTER 4000 OXYGEN TRANSMITTER" on the
preceding page.
1. Install the Oxymitter 4000 in an appropriate location on the stack or
duct. Refer to "Selecting Location" in Section 2: Installation, for
information on selecting a location for the Oxymitter 4000.
2. If using an SPS 4001B, connect the calibration gasses to the
appropriate fittings on the SPS 4001B manifold.
3. Connect reference air to the Oxymitter 4000 or SPS 4001B, as
applicable.
4. If using an SPS 4001B, make the wiring connections as shown in the
SPS 4001B Single Probe Autocalibration Sequencer Instruction
Manual.
5. If NOT using an SPS 4001B, make the following wire connections as
shown in Figure 3: line voltage, 4-20 mA, and logic I/O.
6. Verify the Oxymitter 4000 switch configuration is as desired. Refer to
"Oxymitter 4000 Configuration", "SW1 Setting", and "SW2 Setting" all in
Section 3: Configuration of Oxymitter 4000 with Membrane Keypad, or
"Oxymitter 4000 Configuration", "SW1 Setting", and "SW2 Setting" all in
Section 4: Configuration of Oxymitter 4000 with LOI.
7. Apply power to the Oxymitter 4000; the cell heater will turn on. Allow
approximately one half hour for the cell to heat to operating
temperature. Once the ramp cycle has completed and the
Oxymitter 4000 is at normal operation, proceed with step 8 or 9.
8. If using an SPS 4001B, initiate a semi-automatic calibration.
9. If NOT using an SPS 4001B, perform a manual calibration. Refer to
"Calibration with Keypad" or "Calibration with LOI" both in Section 9:
Maintenance and Service, in this instruction manual.
NOTE
If your system has a membrane keypad you can refer to the Quick Start Guide
on the following pages.