Emerson Process Management 8732 Oxygen Equipment User Manual


 
Reference Manual
00809-0100-4663, Rev BA
January 2010
A-7
Rosemount 8732
Stability
±0.1% of rate over six months
Ambient Temperature Effect
±0.25% change over operating temperature range
EMC Compliance
EN61326-1 1997 + A1/A2/A3 (Industrial) electromagnetic compatibility (EMC)
for process and laboratory apparatus.
PHYSICAL
SPECIFICATIONS
Materials of Construction
Housing
Low copper aluminum, NEMA 4X and IEC 60529 IP66
Pollution Degree 2
Paint
Polyurethane
Cover Gasket
Rubber
Electrical Connections
Two
1
/2–14 NPT connections provided on the transmitter housing (optional
third connection available). PG13.5 and CM20 adapters are available. Screw
terminals provided for all connections. Power wiring connected to transmitter
only. Integrally mounted transmitters are factory wired to the sensor.
Transmitter Weight
Approximately 7 pounds (3.2 kg). Add 1 pound (0.5 kg) for Option Code M5.