Emerson Process Management 8732 Oxygen Equipment User Manual


 
Reference Manual
00809-0100-4663, Rev BA
January 2010
3-3
Rosemount 8732
Table Value Example Setting the TUBE SIZE:
1. Press the DOWN arrow to access the menu.
2. Select line size from the Basic set-up menu.
3. Press the UP or DOWN arrow to increase/decrease (incrementally)
the tube size to the next value.
4. When you reach the desired size, press “E” (the left arrow).
5. Set the loop to manual if necessary, and press “E” again.
After a moment, the LCD will display the new tube size and the maximum flow
rate.
Select Value Example Changing the ANALOG OUTPUT RANGE:
1. Press the DOWN arrow to access the menu.
2. Using the arrow keys, select PV URV from the Basic Setup menu.
3. Press RIGHT arrow key to position the cursor.
4. Press UP or DOWN to set the number.
5. Repeat steps 2 and 3 until desired number is displayed.
6. Press “E”.
After a moment, the LCD will display the new analog output range.
Display Lock The display can be locked to prevent unintentional configuration changes.
The display lock can be activated through a HART communication device, or
by holding the UP arrow for 10 seconds. When the display lock is activated,
DL will appear in the lower left hand corner of the display. To deactivate the
display lock (DL), hold the UP arrow for 10 seconds. Once deactivated, the
DL will no longer appear in the lower left hand corner of the display.
Start Totalizer To start the totalizer, press the DOWN arrow to display the totalizer screen
and press “E” to begin totalization. A symbol will flash in the lower right
hand corner indicating that the meter is totalizing.
Stop Totalizer To stop the totalizer, press the DOWN arrow to display the totalizer screen
and press “E” to end totalization. The flashing symbol will no longer display
in the lower right hand corner indicating that the meter has stopped totalizing.
Reset Totalizer To reset the totalizer, press the DOWN arrow to display the totalizer screen
and follow the procedure above to stop totalization. Once totalization has
stopped, press the RIGHT arrow key to reset the NET total value to zero.
To reset the gross total value, you must change the line size. See “Line Size”
on page 3-9 for details on how to change the line size.