Emerson Process Management 8732 Oxygen Equipment User Manual


 
Reference Manual
00809-0100-4663, Rev BA
January 2010
E-19
Rosemount 8732
KENT VERIFLUX VTC
SENSOR
Connect coil drive and electrode cables as shown in Figure E-16.
Veriflux VTC Sensor to
8732 Transmitter
Figure E-16. Wiring Diagram for
Kent Veriflux VTC Sensor and
Rosemount 8732
Table E-17. Kent Veriflux VTC
Sensor Wiring Connections
Electrode Connections
Coil Connections
ROSEMOUNT 8732
TRANSMITTER
KENT VERIFLUX VTC
SENSOR
1SCR OUT
2
3 SIG 1
4 SIG 2
5
6
1
2–
5+
6 SCR OUT
Fuse
Rosemount 8732 Kent Veriflux VTC Sensors
12
2 1
SCR OUT
17 SCR OUT
18 SIG1
19 SIG2
This is a pulsed DC magnetic flowmeter.
Do not connect AC power to the sensor
or to terminals 1 and 2 of the
transmitter, or replacement of the
electronics board will be necessary.