Set the test criteria for the Empty Pipe condition. The factory default for this
value is set to three percent with limits configurable between one and ten
percent.
8714i Meter Verification Test Scope
The 8714i Meter Verification can be used to verify the entire flowmeter
installation, or individual parts such as the transmitter or sensor. This
parameter is set at the time that the 8714i Meter Verification test is initiated.
All
Run the 8714i Meter Verification test and verify the entire flowmeter
installation. This parameter results in the verification test performing the
transmitter calibration verification, sensor calibration verification, coil health
check, and electrode health check. Transmitter calibration and sensor
calibration are verified to the percentage associated with the test condition
selected when the test was initiated.
Transmitter
Run the 8714i Meter Verification test on the transmitter only. This results in
the verification test only checking the transmitter calibration to the limits of the
test criteria selected when the 8714i Meter Verification test was initiated.
Sensor
Run the 8714i Meter Verification test on the sensor only. This results in the
verification test checking the sensor calibration to the limits of the test criteria
selected when the 8714i Meter Verification test was initiated, verifying the coil
circuit health, and the electrode circuit health.
8714i Meter Verification
Test Results Parameters
Once the 8714i Meter Verification test is initiated, the transmitter will make
several measurements to verify the transmitter calibration, sensor calibration,
coil circuit health, and electrode circuit health. The results of these tests can
be reviewed and recorded on the calibration verification report found on
page C-16. This report can be used to validate that the meter is within the
required calibration limits to comply with governmental regulatory agencies
such as the Environmental Protection Agency or Food and Drug
Administration.
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Set Pass/Fail Criteria, Empty Pipe Limit
AMS Tab 8714i
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Run 8714i
AMS Tab Context Menu, Diagnostics and Tests, 8714i Meter Verification
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Run 8714i
AMS Tab Context Menu, Diagnostics and Tests, 8714i Meter Verification
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Run 8714i
AMS Tab Context Menu, Diagnostics and Tests, 8714i Meter Verification