Emerson Process Management 8732 Oxygen Equipment User Manual


 
Reference Manual
00809-0100-4663, Rev BA
January 2010
C-11
Rosemount 8732
Viewing the 8714i Meter Verification Results
Depending on the method used to view the results, they will be displayed in
either a menu structure, as a method, or in the report format. When using the
375 Field Communicator, each individual component can be viewed as a
menu item. In AMS, the calibration report is populated with the necessary
data eliminating the need to manually complete the report found on
page C-16.
NOTE
When using AMS there are two possible methods that can be used to print the
report. Method one involves taking a screen capture of the 8714i Report tab.
Using Ctrl + Alt + PrntScrn will capture the active window and allow for
pasting of the report directly into a word processing program.
Method two involves using the print feature within AMS while on the status
screen. This will result in a printout of all of the information stored on the
status tabs. Page two of the report will contain all of the necessary calibration
verification result data.
The results are displayed in the following order:
Test Condition
Review the test condition that the 8714i Meter Verification test was performed
under.
Test Criteria
Review the test criteria used to determine the results of the 8714i Meter
Verification tests.
8714i Result
Displays the overall result of the 8714i Meter Verification test as either a Pass
or Fail.
Simulated Velocity
Displays the simulated velocity used to verify the transmitter calibration.
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results, Test Condition
AMS Tab Context Menu, Device Diagnostics, 8714i Report
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results, Test Criteria
AMS Tab Context Menu, Device Diagnostics, 8714i Report
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results, 8714i Result
AMS Tab Context Menu, Device Diagnostics, 8714i Report
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
8714i Results, Simulated Vel
AMS Tab Context Menu, Device Diagnostics, 8714i Report