Emerson Process Management 8732 Oxygen Equipment User Manual


 
Reference Manual
00809-0100-4663, Rev BA
January 2010
4-3
Rosemount 8732
EP Trigger Level
Limits: 3 to 2000
Configure the threshold limit that the empty pipe value must exceed before
the diagnostic alert activates. Default from the factory is set to 100. For more
detail on this parameter see Appendix C: Diagnostics.
EP Counts
Limits: 5 to 50
Configure the number of consecutive times that the empty pipe value must
exceed the empty pipe trigger level before the diagnostic alert activates.
Counts are taken at 1.5 second intervals. Default from the factory is set to 5.
For more detail on this parameter see Appendix C: Diagnostics.
Electronics Temp Value
Electronics Temperature allows you to view the current value for the
electronics temperature.
Advanced Diagnostics The advanced diagnostics menu contains information on all of the additional
diagnostics and tests that are available in the 8732 transmitter if one of the
diagnostics suite packages was ordered.
Rosemount offers two advanced diagnostic suites. Functionality under this
menu will depend on which of these suites are ordered.
Advanced diagnostics suite option one (D01 option) contains advanced
diagnostics for High Process Noise detection and Grounding and Wiring fault
detection.
Advanced diagnostics suite option two (D02 option) contains advanced
diagnostics for the 8714i Meter Verification. This diagnostic is used to verify
the accuracy and performance of the magnetic flow meter installation.
8714i Meter Verification
This diagnostic allows you to test and verify that the sensor, transmitter, or
both are working within specifications. For more details on this diagnostic, see
Appendix C: Diagnostics.
Run 8714i
Run the meter verification test to check the transmitter, sensor, or entire
installation.
Full Meter Verification
Run the internal meter verification to check the entire installation, sensor and
transmitter at the same time.
Transmitter Only
Run the internal meter verification to check the transmitter only.
Sensor Only
Run the internal meter verification to check the sensor only.
375 Transducer Block, Diagnostics, Basic Diagnostics, Empty Pipe Limits
375 Transducer Block, Diagnostics, Basic Diagnostics, Empty Pipe Limits
375 Transducer Block, Diagnostics, Basic Diagnostics
375 Transducer Block, Diagnostics
375 Transducer Block, Diagnostics, Advanced Diagnostics
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification