Emerson Process Management 8732 Oxygen Equipment User Manual


 
Reference Manual
00809-0100-4663, Rev BA
January 2010
Rosemount 8732
4-6
Re-Signature Meter
Have the transmitter measure and store the sensor signature values. These
values will then be used as the baseline for the meter verification test. Use
this when connecting to older Rosemount or competitors’ sensors or installing
the magnetic flowmeter system for the first time. For more details on this
parameter see Appendix C: Diagnostics.
Recall Last Saved Values
Recalls the last saved “good” values for the sensor signature.
Set Pass/Fail Criteria
Set the maximum allowable deviation percentage test criteria for the 8714i
Meter Verification test. There are three tests that this criteria can be set for:
Full Pipe; No Flow (Best test condition) – Default is 2%
Full Pipe; Flowing – Default is 3%
Empty Pipe – Default is 5%
NOTE
If the 8714i Meter Verification test is done with an empty pipe, the electrode
circuit will NOT be tested.
No Flow Limit
Limits: 1 to 10 percent
Set the pass/fail test criteria for the 8714i Meter Verification test at Full Pipe,
No Flow conditions.
Flowing Limit
Limits: 1 to 10 percent
Set the pass/fail test criteria for the 8714i Meter Verification test at Full Pipe,
Flowing conditions.
Empty Pipe Limit
Limits: 1 to 10 percent
Set the pass/fail test criteria for the 8714i Meter Verification test at Empty Pipe
conditions.
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Sensor Signature
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Sensor Signature
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Set Pass/Fail Criteria
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Set Pass/Fail Criteria
375 Transducer Block, Diagnostics, Advanced Diagnostics, 8714i Meter Verification,
Set Pass/Fail Criteria