Emerson Process Management 8732 Oxygen Equipment User Manual


 
Reference Manual
00809-0100-4663, Rev BA
January 2010
A-5
Rosemount 8732
8732E transmitters and other manufacturer’s sensors can be calibrated at
known process conditions or at the Rosemount NIST-Traceable Flow Facility.
Transmitters calibrated on site require a two-step procedure to match a
known flow rate. This procedure can be found in “Universal Trim” on
page 4-11.
Diagnostics
Basic
Self test
Transmitter faults
Tunable empty pipe
Reverse flow
Coil circuit fault
Electronics temperature
Advanced (D01 Suite)
Ground/wiring fault
High process noise
Advanced (D02 Suite)
8714i Meter Verification
PERFORMANCE
SPECIFICATIONS
(System specifications are given using the frequency output and with the unit
at reference conditions.)
Accuracy
Includes the combined effects of linearity, hysteresis, repeatability, and
calibration uncertainty.
Rosemount 8732E with 8705/8707 Sensor:
Standard system accuracy is ±0.25% of rate ±1.0 mm/sec from 0.04 to 6
ft/s (0.01 to 2 m/s); above 6 ft/s (2 m/s), the system has an accuracy of
±0.25% of rate ±1.5 mm/sec.
Optional high accuracy is ±0.15% of rate ±1.0 mm/sec from 0.04 to 13 ft/s
(0.01 to 4 m/s); above 13 ft/s (4 m/s), the system has an accuracy of
±0.18% of rate.
(1)
Rosemount 8732E with 8711 Sensor:
Standard system accuracy is ±0.25% of rate ±2.0 mm/sec from 0.04 to 39
ft/s (0.01 to 12 m/s).
(1) For Sensor sizes greater than 12 in. (300 mm) the high accuracy is ±0.25% of rate from 3 to
39 ft/sec (1 to 12 m/sec).