Brother BAS-311G Wheelchair User Manual


 
9. TABLE OF ERROR CODES
BAS-311G, BAS-326G
139
Communication and memory-related errors
Code
Cause and remedy
E401
Connection communication error with power supply motor P. C. board detected when power was turned on.
Turn off the power, and then check that connector P6 on the power supply motor P.C. board and connector P17
on the main P.C. board are properly connected.
E402
Communication error with programmer detected when power is turned on.
Turn off the power, and then check that connector P7 on the main P.C. board is properly connected.
E403
Connection error with PMD P. C. board detected when power was turned on.
Turn off the power, and then check that connector P1 on the PMD P.C. board is properly connected.
E410
Communication error with main P. C. board detected.
Turn off the power, and then turn it back on again.
E411
Communication error with power supply motor P. C. board detected.
Turn off the power, and then turn it back on again.
Check if the fuse (F4) is blown.
E412
Communication error with programmer detected.
Turn off the power, and then turn it back on again.
E413
Communication error with PMD P. C. board detected.
Turn off the power, and then turn it back on again.
E420
No CF card is inserted.
E421
Program contents are incorrect and cannot be used, or no data.
Check that the CF card or floppy disk contains data with this program number.
E422
Error occurred while reading the CF card or floppy disk.
Check the data on the CF card or floppy disk.
E424
Insufficient free space on CF card.
Use a different CF card.
E425
Error occurred while writing to the CF card or floppy disk.
Use the specified type of CF card.
Check if the floppy disk is write-protected and if it has enough free space.
E426
R/W key has not been pressed.
Press the R/W key to load the data.
E427
CF card or floppy disk does not contain any data with this program number.
Check the program number.
Data that has already been loaded into internal memory can be used without having to press the R/W key again.
E430
Data cannot be backed up to main P.C. board.
Turn off the power, and then turn it back on again.
E440
Data memory error on main P.C. board.
Turn off the power, and then turn it back on again.
E450
Model selection cannot be read from the machine head memory.
Turn off the power and check that connector P3 on the power supply motor P.C. board is properly connected.
E451
Data cannot be backed up to machine head memory.
Turn off the power, and then turn it back on again.
E452
Machine head memory is not connected.
Turn off the power and check that connector P3 on the power supply motor P.C. board is properly connected.
E474
Internal memory is full and copying is not possible.
Clear the sewing data, or change the settings so that sewing data is not copied to internal memory when it is
loaded (memory switch No. 466 = ON).