2.INTRODUCTION
2.19060ZTRANSMITTER
TheTeledyneAnalyticalInstruments9060Zoxygentransmitterisdesignedtomeasuretheoxygengas
concentration in combustion applications such as power stations and industrial boilers. The
measurement isused to control theair / fuelration inthe burnerto maintain efficiency andsafe
operationconditions.
ThetransmitterisdesignedtobeusedinconjunctionwithaTeledyneAnalyticalInstrumentsprobe
model9060Hor 9060UL/UH.These twoprobesusea zirconiasensorthat mustoperateat>650°C.
The9060Hprobehasanintegratedheater,andthe9060UL/UHprobeusestheprocessheattoheat
the sensor. There is a unique cable connection between th
e 9060Z transmitter and the
9060H/9060UL/UHprobe.
Themodel9060Zoxygentransmitterisbasedonthewellknownmodel9060Htransmitter.Itincludes
anumberofhardwareandsoftwareimprovementssuchasagraphicdisplay,largercharacters,faster
microprocessor, simplified set up menu, alarm logging, faster probe heater control and more
calculat
edvaluessuchasburnerefficiency.
The9060Z OxygenTransmitter hasavariety ofuserselectablefunctions. Theyare simpletouse
becauseeachselectionismenudriven.Foroptionsyouarenotsureabout,readthemanualonthat
particulariteminchapter6,SetupMode.
Featuresinclude:‐
Inputs
Twozirconiaoxygenprobes,heatedorunheated
Oxygen
read01%to0100%
Furnace,kilnorfluethermocouple,fieldselectableastypeK,J,RorS
Mainflameestablishedsafetyinterlock(forheatedprobesonly)
Purgepressureorflowswitch
Outputs
Twolinearised420mAor020mADCisolatedoutputs,max.load1000Ω
Theoutputfunctionandtherangearefieldselectable
Commonalarmrel
ay
Threeotheralarmrelayswithselectablefunctions
DigitalInterface
RS232ortwowireRS485MODBUSforconnectiontoacomputer/DCS/PCLfordiagnosticsofthe
transmitter,probeorcombustionprocess.
Display
Multifontgraphicaldisplay
Largefontcharactersfortheoxygenonthetopline
Selectablelowerlineitemsforthesecondarydisplayfunctions.ieProbetemperature,Oxygensecond
probeAlarmdisplaymodethatshowsthetimeofthealarm,theacceptancetimeandthetimethat
thealarmwasclearedofupto4000alarmevents
2.2SERIES9060HOXYGENPROBES&SENSORS
Pleasenotethatifthisequipmentisnotinstalledandusedinthemannerdescribedinthismanualthen
thesafetyprotectionprovidedbytheequipmentmaybeimpaired.