Teledyne 9060Z Oxygen Equipment User Manual


 
Model9060ZManual
51

11.GASCALIBRATIONCHECKANDPURGE
Theoxygen sensorthatisusedintheTeledyneAnalyticalInstrumentsoxygen probeisextremely
predictable,stableandreliable.Forthisreason,thecalibrationofaTeledyneAnalyticalInstruments
oxygen system does not require the use of calibration gases. However, all Teledyne Analytical
Instrumentsoxygenprobeshaveabuiltingasconnectionthatdoesallowtheaccuracyoftheprobeto
becheckedinsitu.Thischapterdescribestheoperationofthisgascheckingsystem.
The 9060Z has a timer and solenoid driving system that can be configured to admit a certified
calibrationgasintotheprobeoranairsupplytopurgetheprobefiltersthroughthegasconnection.
Boththecalibrationgasandthefilterpurgegasmustbepipedtotheportontheprobelabeled“CAL”.
Therearetwosolenoidsdriversinthe9060Ztransmitter.Theycanbeusedforanycombinationsof
gascheckingand probe filter purging.The only limitationis thatif the transmitter isin dual probe
mode,thensolenoid#1mustbewiredandplumbedtotheprobe#1andsolenoid#2mustbewired
andplumbedtoprobe#2.
ThetransmittercanalsobeconfiguredtobeinaMANUALorAUTOMATICpurgeandgascheckmode.
Function# FunctionDescription Options
20(29) Solenoid1(2)Operation CalibrationGas/Purge √ √ √
21(30) Solenoid1(2)Auto/Manual Automatic/Manual  √ √ √
22(31) Solenoid1(2)StartTime 00:00to23:45√
23(32) Solenoid1(2)Period 1hour7days√
24(33) Solenoid1(2)Duration 190seconds√
25(34) Solenoid1(2)PostFreeze 5300seconds√
26(35) OxygenContentCalGas1(2) 0.120.9%
27(36) MaxCalGas1(2)PositiveError 0.1%to3.0%√
28(37) MaxCalGas1(2)NegativeError 0.1%to3.0%√
11.1PURGE
Forprocessesthathaveparticulatepresentinthegasstream,probefiltersarenecessaryandshould
bebackpurgedwithsufficientfrequencytoavoidblockedfilters.Select‘Purge’infunction#20(29)
and‘Automatic’infunction#30andthensetthetimesinfunction#22(31)to#25(34)Theoutputs
canbefrozenduringpurgingifselectedinCOMMISSIONINGmenufunction#19.Aflowswitchcanbe
wiredtoterminals16and17.Iftheflowisnotsufficientattheendofthepurgecycletoclosethe
contactsoftheflowswitcha‘Probe 1(2)FilterBlocked’ alarmisgenerated.If aflowswitchisnot
beingusedashortcircuitshouldbeplacedacrossterminals16and17tosuppressthe‘Probe1(2)
FilterBlocked’alarm.
11.2CALIBRATIONGAS
Ifitisrequiredtoautomaticallychecktheoxygenprobeandtransmitteronline,select‘Calibration
Gas’infunction#20(29)and‘Automatic’infunction#30andthensetthetimesinfunction#22(31)
to#25(34)andthegaslevelsinfunction#26(35)to28(37).Ifthetransmittermeasurementdoesnot
fall within the limits set in function#26(35) to28 (37)a ‘Gas1 (2) Calibration Error’ alarm will be
generated.Ifthetransmitterhas‘Automatic’selectedinfunction#21(30)andoneoftheGAS/PURGE
keysispressedinRUNmode,theautomaticcyclewillbestartedandth
ealarmswillbechecked.The
cyclecanbestoppedatanytimebypressingthesamekeyduringthecycle.
FunctionsavailableifeitherCalibrationGasorPurge andManualare selected
FunctionsavailableifPur
g
e andAutomaticareselected
FunctionsavailableifCalibrationGas andAutomaticareselected