9.3ALARMRELAYOPTIONS
Thethreeprocessalarmrelays(relays1to3)areuserdefined.Therelayactionwilldependonthe
selectionsmadeintheCOMMISSIONINGmenufunctions#45to#48.
Therelaycontactsarealldesignedtobe“FailSafe”.Thatis,theyare
openwhenthetransmitterpowerisoff
closedwhenthereisNOTanalarmcondition
openwhenanalarmoccurs
Thecontactwillclosecircuitagain(relayenergised)whenthealarmconditionisacknowledged.
Choosetohavetherelaysreacttoanyorallofthefollowingalarmsorwarnings
Oxygenonprobe1Low Alarm
Oxygenonprobe2Low Alarm
Oxygenonprobe1VeryLow Alarm
Oxygenonprobe2VeryLow Alarm
Oxygenonprobe1High Alarm
Oxygenonprobe2High Alarm
OxygenDeviationHigh Alarm
Probe1TemperatureLow Warning
Probe2TemperatureLow Warning
Calibrationcheckonprobe1in
Progress
Warning
Calibrationcheckonprobe2in
Progress
Warning
Purgeonprobe1inProgress Warning
Purgeonprobe2inProgress Warning
Inadditiontotheaboveprocessalarmsthatcanactivatetheprocessalarmrelays,anyofthecommon
alarm relay events that have been taken off the common alarm list in COMMISSIONING menu
function#49willappearontheprocessalarmrelaylistsinfunction#46to#48.
