7.2.25CommunicationsPortOptions
The9060Ztransmitterhasaserialcommunicationsportavailableatterminals18to22.Thedefault
protocolisforRS‐232runningat19,200baudratewith8bits,Evenparityand1stopbit.
Function Options
60 SerialInterface RS‐232orRS‐485(UseRS‐485forMODBUS)
61 SerialBaudRate 9600,19200,38400,57600(RS‐232upto19200only)
62 SerialParity Even,OddorNoParity
63
MODBUS
Address
SettheMODBUSaddressforthisdevicebetween1and246.
0istodisabletheMODBUS
7.2.26AlarmLogClearing
Everyalarmthatisinstigatedinternallyinthetransmitterorasaprocesslevelisrecordedinthealarm
log with the activation time, accepted time and the cleared time. The last 4000 events will be
recordedandthentheoldestalarmswillbedroppedoffasnewonesoccur.Ifyouwo
uldliketodelete
thelogrecordingselectCLEARinSETUPstep64byusingtheOPTIONkeysandpresstheENTERkey.
Themessage“AlarmLogCleared”willbedisplayed.