Teledyne 9060Z Oxygen Equipment User Manual


 
Model9060ZManual
23

Theonlineauto purgeand calibrationchecksystem isoptional. Fordetailsonits operationsee
chapters4.24,4.25andchapter11.
Toautomaticallysenseablockedprobefilter,aflowswitchshouldbeinstalledinthe‘purge’lineto
theprobe‘CAL’port.Itshouldbeadjustedsothatitenergizesjustbelowthepurgeflowwithnewor
cleanfiltersinstalled.Theflowswitchcontactsshouldbeconnectedtoterminals16&17(FLOWSW).
Ifthefilterisstillblockedorpartlyblockedafteranautopurgecycle,theflowswitchwillnotenergize
andwillthereforecausea‘Probe1(2)FilterBlocked’alarm.
Afterinstallationthe purge/calsystem should be testedthoroughly forleaks. Anyleaks cancause
significanterrorsiftheflueisatnegativepressure.Iftheflueisatpositivepressure,anoutwardleak
cancausecorrosioninthepurge/calsystempipingandfittings.
Ifprobe/filterpurgingisrequiredbuta“Probe1(2)FilterBlocked”alarmisnotrequired,donotinstall
theflowswitchbutlinkterminals16&17.
4.12CONNECTINGREFERENCEAIR
Foroxygenprobes,a1/4”tubeconnectoronthetransmittershouldbeconnectedviaanylon,copper
orstainlesssteeltubetothe‘REF’connectorontheprobe.Iftwoprobesarebeingused,a“T”fitting
mustbesuppliedtoprovidereferenceairsupplytobothprobes.
4.13CONNECTINGTHETRANSMITTE R TOAMODBUS™NETWORK
Thetransmittercanbenetworkedtoothertransmittersandtoanetworkmaster.Thenetworkuses
the transmitter’s RS485port. Up to31transmitters can beconnectedtothe network, and canbe
interrogatedbytheNetworkMaster.
NOTE:Hardware Protocol SelectionThe communicationsport of the9060Z canbe configuredto
communicateoneitherRS232orRS485.IfthetransmitteristobeusedonaMODBUSnetwork,the
9060ZtransmittermustbesettoRS485.Forfurtherdetailsseechapter7.2.25,CommunicationPort
Options.
NOTE:TerminatingResistorThereisaterminating100ohmresistorfittedtothe17321PCB.Lin
kLK3
(TERM),inthebottomlefthandcornerofthePCB,isusedtoconnecttheterminatingresistor.Link
LK3mustberemovedonalltransmittersexceptthetransmitterontheendofthenetworkline.Ifthe
AutomaticPurge&
Calibration check
System Wiring
Schematic
AutomaticPurge&
Calibration check
System Piping
Schematic