6.3.3OxygenDisplayUnits
Thetoplineofthedisplaythatshowstheoxygenmeasurementcanbedisplayedineither%orppm
(seechapter5.2OxygenDisplayUnits)
6.3.4DampingFactor
Theoxygenmeasurementcanbedampedbyaveragingsuccessivereadingsfromtheprobe.Thiswill
smooth out fluctuations in the process gas level and will slow down the reaction time of the
transmitter.Thelargerthenumberselectedhere,themoresuccessivereadingsareaveragedandthe
smootherthe measurementwillbe. The dampedoxygenvalue isalsousedin thecalculationofall
otherparametersthatarebasedontheoxygenvalue.
6.3.5ProcessAlarms
Thisfunctionallowstheoperatortoenableanddisableprocessalarms.Theprocessalarmmenu
functionswill alsobehidden iftheprocess alarmsaredisabled.(Seealso chapter7.2.19,Process
Alarms)