1. IMPORTANTNOTICES
Thismanualisintendedasasupplementtothe9060ZOperatorsManual.Itisintendedtobeusedby
technicalpersonalthatarequalifiedtoinstall,commission,serviceandcalibrateelectronicindustrial
controlequipment.
Pleasereadthesafetyinformationbelowbeforeconnectingpowertothetransmitter.
CAUTION1
The probe heater is supplied with mains voltage. This supply has electrical shock danger to
maintenance personnel. Always isolate the transmitter before working withthe probe. The EARTH
wires(green/yellow)fromaheatedprobemustALWAYSbeconnectedtoearth.
CAUTION2
Combustionoratmospherecontrolsystemscanbedangerous.Burnersmustbemechanicallysetup
so that in the case of equipment failure the system cannot generate explosive atmospheres. This
dangerisnormallyavoidedwithfluegastrimsystemsbyadjustmentsothatinthecaseoffailurethe
appliancewillnotgenerateCOinexcessof400ppmintheflue.TheCOlevelintheflueshouldbe
measuredwithaseparateCOinstrumentnormallyaninfraredorfuelcelltype.
CAUTION3
Theoxygenprobeisheatedtoover700°C(1300°F)andisasourceofignition.Sincerawfuelleakscan
occurduringburnershutdown,thetransmitterhasaninterlockingrelaythatremovespowerfromthe
probeheaterwhenthemainfuelshut‐offvalvepowerisoff.Ifthisconfigurationdoesnotsuitorifit
is possible for raw fuel to come into contact with a hot oxygen probe then the Model 9060Z
transmitter with a heated probe will not be safe in your application. An unheated probe can be
utilizedinsuchapplications,howevertheoxygenreadingsarevalidonlyabove650°C(1200°F).
CAUTION4
Thereducingoxygensignalfromthetransmitterandtheassociatedalarmrelaycanbeusedasan
explosive warning or trip. This measurement assumes complete combustion. If incomplete
combustionispossiblethenthissignalwillreadlessreducingandshouldnotbeusedasanalarmor
trip.Atrueexcesscombustiblesanalyzer,normallyincorporatingacatalystorthermalconductivity
bridge,wouldbemoreappropriatewhereincompletecombustionispossible.Alsoreadtheprobe
electricalshockcautionintheprobeheaterinterlockcautioninchapter4.7
CAUTION5
FIL‐3filter.IftheoptionalFIL‐3hasbeenfittedtothe9060Hprobeinthisinstallation,pleasereadthe
ImportantNoticeregardingprobeoptionFIL‐3onthepreviouspage.
CAUTION6
CAUTION7
The heater is supplied from the mains power directly, and the temperature is controlled at 720°C
(1330°F).Theoutsideoftheprocessendoftheprobecangettotemperaturesthataredangerousto
touch.Wearinsulatinggloveswhenhandlingaprobethathasbeenon.