Teledyne 9060Z Oxygen Equipment User Manual


 
Model9060ZManual
15

Theoxygendeficiencyoutputcanbeusedinthesamewayasacombustiblesanalyzertosignalthe
extentofreducingconditionsofcombustionprocesses.
4.INSTALLATIONANDCOMMISSIONING
4.1MOUNTINGTHETRANSMITTER
Surfacemountthetransmittercaseontoaflatsurfaceorbracket,usingthefourmountingbrackets
provided.Thetransmittershould never bemountedsothatit isdirectlyexposeto thesunorrain.
Alwaysleaveatleast10cmofclearancearoundthefoursidesofthecase.
Makesurethetemperatureofth
ecaseisbelow55°C,andthattheradiatedheatfromfurnacesand
boilersiskepttoaminimum.Thereshouldbeadequateventilationtomaintainambienttemperature.
Installthecablesthroughcableglands.Thereare4holescutinthebaseofthetransmittercase;2x
17mm&2x21mm.Useasharpknifetocutawaythecoveringfilmforonlytheglandsthatare
needed.
NOTE:ALWAYSLEAVETHEUNUSEDGLANDHOLESSEALED.
Ensurethattheelectricalconnectioncomplieswiththelocalelectricalrequirements.(seechapter4.6,
ElectricalConnections)
4.2INSTALLINGA9060HOXYGENPROBE
WeldaBSPorNPTsockettotheflueinasuitablepositionforfluegassensing.Forthecorrectsizeof
socketrefertoprobedatainchapter3.1.3.Theclosertothesourceofcombustionthesmallerwillbe
sensinglagtime,allowingbettercontrol.
Theprobehasatypicalresponsetimeoflessthanfourseconds,somostofthedelaytimeisnormally
thetransittimeofthegasfromthepointofcombustiontothepointofsensing.