6.SETUPMODE
The SETUP mode ismode is accessed by pressingtheSETUP keymomentarily so the words ‘Setup
Mode’appearatthebottomofthedisplay.ThetransmitterwillreturntotheRUNmodewhenthe
SETUPkeyispressedagainor60secondsafterthelastkeypress.
6.1FUNCTIONSUMMARYTABLE
WhenthetransmitterisintheSETUPmodetheSETUPlightwillbeonandthewords“SetupMenu”
willbeshownatthebottomofthedisplay.
ThefollowingtableshowstheSETUPfunctions:
Menu# Functionname Range Defaultvalue
1 Probe1Offset 6.0to+6.0mV 0.0mV
2 Probe2Offset 6.0to+6.0mV 0.0mV
3 LowerLineItems SeeChapter6.3.2
4 OxygenDisplayUnits Oxygen%/Oxygenppm Oxygen%
5 DampingFactor NoDampingto10Samples 5Samples
6 ProcessAlarms Enabled/Disabled Disabled
6.2CHANGINGOPTIONS
The purposeof having a menudrivenSETUP modeis to allow forconfigurationof the transmitter
usingthegraphicaldisplayandkeypad.
OnceanoptionischangedandenteredusingtheENTERkeythatvalueimmediatelybecomesactive.
ItisalsostoredintothenonvolatilememorywithintheBBRAMandwillbeloadedagainatdevice
startup.Toresetthetransmitterconfigurationdefaultsseechapter10.2,ColdStart.
TochangeanoptionintheSETUPmenusystem:
1. SelecttheSETUPmodebypressingtheSETUP/RUNkeyonce.TheSETUPlightwillcomeonand
thedisplaywillhaveth
eformatshownbelow.Theoperationsofthekeysarenowtheoperations
writteninWHITEonthekeypad.Themenunameiswrittenatthebottomofthedisplay.
2. When the SETUP mode has been selected the required function can be found by using the
FUNCTIONUPandFUNCTIONDOWNkeys.
TheoptionsavailableforthatfunctioncanbeseenbyusingtheOPTIONUP&OPTIONDOWN
keys.
3. WhentherequiredoptionisonthedisplaypresstheENTERkeytosavethatvalue.
Whenfinished,presstheSETUP/RUNkeytoreturntotheRUNmode.