5.2OXYGENDISPLAYUNITS
Themeasured oxygencontentdisplayed on the top lineofthetransmitter for probe 1and onthe
lowerlineforprobe2canbeshownineither%orpartspermillion(ppm).Seechapter6.3.3Oxygen
DisplayUnitsforhowtochangetheoxygendisplay.
Theformatoftheoxygendisplaychangestomaintainthebestresolutionforthemeasurement.
If‘Oxygen%’isselectedthetransmitterwilldisplayoxygeninthefollowingformat:
RangeDisplayformat
30.0to100.0% ##.#%
1.00to29.99% ##.##%
0.100to0.999% #.###%
<0.100% scientificnotation(#.##x10‐##%)
If‘Oxygenppm’ isselectedthetransmitterwill showtheoxygenin ppmformatbetween10 and
9,999ppm.Outsideofthisrangetheoxygenwillbeshownin%formatdescribedabove.
Displayformatfor‘Oxygenppm’:
RangeDisplayformat
>10,000ppm(1%) %format(see‘Oxygen%’sectionabove)
1,000to9,999ppm #,###ppm
100.0to999.9ppm ###.#ppm
10.00to99.99ppm ##.##ppm
<10ppm scientificnotation(#.##x10‐##%)
5.3KEYPAD
Thereareeightkeysbuiltintothedecalontheoutsideofthedoorofthe9060Ztransmitter.Thekey
functionsarewritteninBLACKandWHITEtoidentifythefunctionofthekeyineitherRUNmodeor
SETUPmode.
Keytext RUNmode(WHITEtext) SETUPmode(BLACKtext)
SETUP/RUN EnterSETUPmode ReturntoRUNmode
DISPLAY/FUNCTIONUP DisplayNextLowerLineItem NextFunction
DISPLAY/FUNCTION
DOWN
DisplayPrev.LowerLineItem
PreviousFunction
ALARM/OPTIONUP NextAlarm Next/IncrementOption
ALARM/OPTIONDOWN PreviousAlarm Previous/DecrementOption
ALARMACCEPT/ENTER AcknowledgeDisplayedAlarm SaveCurrentOption
GAS1PURGE1/SENSIMP Gas1/Purge1manualactivate CheckProbeimpedance
GAS2PURGE2/AUTOCAL
Gas2/Purge2manualactivate
ManuallyActivateOutputCalibration
5.3.1KeypadinRUNmode
Whenthetransmitteristurnedonandhascompleteditsstartupprocedure,itwillenterRUNmode
andenablethekeypad.Thekeyfunctionsinthismodeareasfollows;