highimpedancealarm,‘Probe1(2)HighImpedance’,isnotactive.Theactualimpedancecanbe
displayedonthelowerline.Itshouldbelessthan9KΩat720°C(1320°F).
Onceithasbeenestablishedthattheimpedanceisnormal,theoffsetmaybesetusingthemillivolt
valuemarkedontheoxygenprobe.Seechapter6.3.1,ProbeOffset.Theprobeoffsetcanbetestedon
site.Asmallflowofairmustbeadmittedtoboththe‘REF’and‘CAL’portswhentestingtheprobe
offset.Iftheprobeisintheprocess,theairmustfullypurgetheprobesensorwithoutinterfer
ence
fromtheprocessgassample.TeledyneAnalyticalInstrumentsprobescaneasilyachievethiswithor
withoutaprobefilterandagasflowofonly1to5SLPM(2to10SCFH)fora9060Hprobeandupto
20SLPM(40SCFH)foranunheatedprobe.Whenastableoxygenre
adinghasbeenachieved,readthe
‘ProbeEMF’.Enterthisvalueintothesetupmenu#1fortheprobe1(setup#2forthesecondprobe
ifitisinstalled).
4.17FILTERPURGING
Purgingprobefiltersiscontrolledfromthe‘GAS/PURGE’buttonson thetransmitterwhenin‘RUN’
mode. If ‘Automatic’ has been selected in either ‘Sol. 1 Auto/Man’ or ‘Sol. 2 Auto/Man’ in
Commissioning Menus 21 or 30, pressing the GAS/PURGE button will start the automatic cycle.
Pressing the button again will cancel the auto purge cycle. If ‘Automatic’ was not enabled, the
solenoidwillonlystayopenforaslongasthebuttonispressed.Graduallyadjustthepurgeairsupply
regulator, increasing the pressureuntil sufficient flow is obtained to clear thefilter. This is best
checkedwith a dirty filterafter a period of operation, by withdrawing the probe from service and
watching any build uponthe filter being blown off at the set pressure. Normally 30kPa(5 psi) is
adequatebuttheairpressuremaybesetashighas100kPa(15psi).
4.18CALIBRATIONGASCHECK
Iftheinstallationhasafilterpurgefacility,setthisupfirst.Refertothepreviousparagraph.Pressthe
‘GAS/PURGE1or‘GAS/PURGE2’buttonwhilein‘SETUP’modetoobtainareasonableflowthrough
thecalibrationcheckgasflowmeter.Ifairisbeingusedasacalibrationcheckgas,usetheairfromthe
regulatorforfilterpurge.Then,whensettingupagasforcalibrationchecking,setthepressurefrom
thecalibrationgascylindersothatitisthesameasthepressuresetontheairregulator.Thenthe
settingontherotameter/flowregulatorwillbethesameasthatfortheairflow.Theflowrequiredis
1to5SLPM(2to10SCFH)fora9060Hprobeandupto20SL
PM(40SCFH)foranunheatedprobe.
Airisnotthebestgasforcalibrationcheckingonazirconiasensor.Theoutputofazirconiasensor
withaironbothsidesofthesensoriszeromillivolts.Itisbettertochooseagasvaluewhichprovides
areasonableoutputfromthesensorandwhichisneartotheprocessoxygenlevel.Acylinderwith2%
oxygeninnitrogenisacommonlyusedcalibrationgas.Themaximum pressureonthe calibration
checkgascylinderregulatorsis100kPa(15psi).
Note:If‘Dualprobe’wasselectedincommissioningmenu4,‘CalGas2’mustbeconnectedtoprobe2.
4.19DUSTINTHEFLUEGAS
Forheatedprobesthepreferredmethodofmountingfordustladenapplicationsisfacingvertically
For unheated probes with no filter, entrained solidsor dust in the flue gas does not present a
problemunlessthedust,whensettled,isnotporous.Allowthedustintheprocesstobuildupon
theprobe.Itwillformaporouslayerslowingtheresponsetime.Toavoidmechanicalabrasionof
theelectrodematerialininstallationswithunheatedoxygenprobes,pack‘SAFFIL’orequivalent
aluminabasedceramicfiberinthesensingholestoprotecttheelectrode.Donotusesilicabased
ceramicfiberssuchas‘KAOWOOL’,whichcanattacktheelectrodeathightemperatures.Oncethe
dusthasbuiltu
p
theres
p
onsetimeofthe
p
robewill beslower.