ImportantNoticeRegarding9060 ProbeOption‐FIL3
WARNING:TheonlyidentifiablestandardforflamearrestersforgeneraluseisBritishStandardBS
EN12874:2001.BritishStandardBSEN12874:2001referstoanoperatingenvironmentupto150
DegreesCentigrade.
The FIL3 device optionally fitted to 9060 Heated Zirconia Probes (the “Probes" or "Probe")
operateinanenvironmentconsiderablygreaterthan150DegreesCentigrade.
Therefore, we know of no Australian, British, European or USA standard applicable to flame
arrestersortheirtestingabove150degreesCentigrade.Consequently,theFIL3devicecannotbe
certifiedasasafetydevice.
Theprobeisonlyonprobesduringthestartupprocessesofacombustionappliance.
TheTeledyneAnalyticalInstrumentsBurnerInterlockRelayfacility,whichisastandardpartofthe
TeledyneAnalyticalInstrumentstransmitter,isdesignedtobewiredtothemainsafetyshutoff
fuelvalvesinawaythatcanshutdowntheprobeheaterwhenthefuelvalvesareclosed.
TheriskofignitionofflammablegasmixtureatthehotendoftheProbecanonlybeminimizedby
correct use, maintenance and operation of the FIL3 device. The user of the FIL3 device is
responsibleforverificationandmaintenanceandcorrectuseandoperationoftheFIL3device.
THEUSERAGREESTHATITUSESTHEPROBEANDTHEFIL3DEVICEATITSSOLERISK.TELEDYNE
ANALYTICALINSTRUMENTSTOTHEFULLEXTENTPERMITTEDBYLAW,GIVESNOWARRANTIESOR
ASSURANCESANDEXCLUDESALLLIABILITY(INCLUDINGLIABILITYFORNEGLIGENCE)INRELATION
TOTHEPROBEANDTHEFIL3DE
VICE.
Theuser mustensure thatitcorrectlyfollows allinstructions inrelation tothe ProbeandFIL3
device,correctlyunderstandsthespecificationsoftheProbeandFIL3deviceandensuresthatthe
ProbeandFIL3deviceareregularlyinspectedandmaintained.
FIL3equippedProbessho
uldbeinspectedatleastonceayearforcorrosionandmorefrequently
ifthereisanyreasontosuspectthatcorrosionmayhaveoccurred.