ImportantNoticeRegarding9060 ProbeOption‐FIL‐3
WARNING:TheonlyidentifiablestandardforflamearrestersforgeneraluseisBritishStandardBS
EN12874:2001.BritishStandardBSEN12874:2001referstoanoperatingenvironmentupto150
DegreesCentigrade.
The FIL‐3 device optionally fitted to 9060 Heated Zirconia Probes (the “Probes" or "Probe")
operateinanenvironmentconsiderablygreaterthan150DegreesCentigrade.
Therefore, we know of no Australian, British, European or USA standard applicable to flame
arrestersortheirtestingabove150degreesCentigrade.Consequently,theFIL‐3devicecannotbe
certifiedasasafetydevice.
Theprobeisonlyonprobesduringthestartupprocessesofacombustionappliance.
TheTeledyneAnalyticalInstrumentsBurnerInterlockRelayfacility,whichisastandardpartofthe
TeledyneAnalyticalInstrumentstransmitter,isdesignedtobewiredtothemainsafetyshut‐off
fuelvalvesinawaythatcanshutdowntheprobeheaterwhenthefuelvalvesareclosed.
TheriskofignitionofflammablegasmixtureatthehotendoftheProbecanonlybeminimizedby
correct use, maintenance and operation of the FIL‐3 device. The user of the FIL‐3 device is
responsibleforverificationandmaintenanceandcorrectuseandoperationoftheFIL‐3device.
THEUSERAGREESTHATITUSESTHEPROBEANDTHEFIL‐3DEVICEATITSSOLERISK.TELEDYNE
ANALYTICALINSTRUMENTSTOTHEFULLEXTENTPERMITTEDBYLAW,GIVESNOWARRANTIESOR
ASSURANCESANDEXCLUDESALLLIABILITY(INCLUDINGLIABILITYFORNEGLIGENCE)INRELATION
TOTHEPROBEANDTHEFIL‐3DE
VICE.
Theuser mustensure thatitcorrectlyfollows allinstructions inrelation tothe ProbeandFIL‐3
device,correctlyunderstandsthespecificationsoftheProbeandFIL‐3deviceandensuresthatthe
ProbeandFIL‐3deviceareregularlyinspectedandmaintained.
FIL‐3equippedProbessho
uldbeinspectedatleastonceayearforcorrosionandmorefrequently
ifthereisanyreasontosuspectthatcorrosionmayhaveoccurred.